Goldmann tonometer calibration

Your Goldmann tonometer will come with a calibration rod as per the image below.

  1. Place tonometer on the guide plate (or move into position if slit-lamp arm mounted tonometer).
  2. Ensure tonometer head is in the holder.
  3. Insert calibration rod into the hole on the side of the tonometer.

Troubleshooting tip: Ensure the longest end of the calibration rod  is pointing away from the patient when using the 20mmHg and 60mmHg point 

  1. Start on the ‘0’ mark. Turn the dial 0.5 mmHg downwards/anti-clockwise. The tonometer probe should rock backwards, towards the examiner.
  2. Now turn the dial 0.5mmHg forwards/clockwise. The tonometer probe should rock forwards, towards the patient.
  3. Adjust the calibration rod to the ‘20’ mark and repeat the above process either side of the mark.
  4. Repeat again for the ‘60’ mark

  1. All three positions should be within the margin of error (+/- 0.5mmHg). If the tonometer is outside the margin of error it should be sent for servicing and calibration as it will not read accurately.